Yıl 2018, Cilt 4, Sayı 1, Sayfalar 37 - 40 2018-06-26

CAPILLARY REFILL TIME MEASUREMENT DEVICE
KILCAL DAMAR DOLUM SÜRESİ ÖLÇÜM CİHAZI

Ufuk Bal [1]

44 46

Capillary refill time (CRT) can be defined as a rapid clinical test for assessing blood flow through peripheral tissues. Importance of the monitarisation of the CRT is emphasized in diagnosis and treatment guideleness for shock.  The capillary refill test is currently conducted manually and its implementation varies according to medical personnel.  There isn’t any commercial device that standardizes both the magnitude and the duration of the force applied for sensing the perfusion of the capillaries. A microcontroller based device is designed to measure the capillary refill time. An infrared led is  used as a light source.  Also a linear actuator is used to able to apply a  constant force during a constant time to the fingertip. We evaluated our design on one healthy volunteer and one thalassemia carrier by performing 30 measurements. It is known that thalassemia carriers may have prolonged capillary refill time which is consistent with our measurements obtained by the proposed design.


Kapiler dolum zamanı, çevresel dokulardaki kan akışını değerlediren hızlı klinik bir test olarak tanımlanabilir. Şokun tanı ve tedavi rehberlerinde kapiller geri dolum zamanı (Capillary Refill Time, CRT) monitorizasyonunun önemi belirtilmektedir. Hâlihazırda CRT manuel olarak ölçülmekte ve uygulanması ölçüm yapan personele göre değişmektedir. Kılcal damarlardaki kanlanmayı tespit etmek için uygulanan kuvveti ve baskı süresini standardize eden ticari cihaz mevcut değildir. Kapiler dolum zamanını ölçmek için mikroişlemci tabanlı bir cihaz tasarlanmıştır. Işık kaynağı olarak kızılötesi bir LED kullanılmıştır. Ayrıca parmak ucuna belirli bir süre bilinen bir kuvvet uygulamak için bir lineer aktüatör kullanılmıştır. Tasarımımızı, bir sağlıklı ve bir talesemi taşıyıcısı gönüllü üzerinde 30 ölçüm gerçekleştirmek suretiyle değerlendirdik. Talesemi taşıyıcılarının daha uzun CRT’ye sahip olduğu bilinmektedir. Önerilen tasarımla yapılan ölçümler de bu durumu doğrulamıştır.

  • Bordoley, A., et al., D1GIT: Automated, temperature calibrated measurement of capillary refill time, 2012, The University of Pennsylvania Philadelphia, Pennsylvania, USA.
  • KN, S. and S. PR., Resuscitation-Pediatric Advanced Life Support, in Textbook of Pediatric Emergency Medicine F. GR and L. S, Editors. 2008, Lippincott Williams & Wilkins.
  • RA, D., Pediatric Assessment, in The Pediatric Emergency Medicine Resource APLS 2007, Jones & Bartlett Learning.
  • EY, F. and H. FM, Evaluation of Vital signs, in Textbook of Pediatric Emergency Procedures, K. C and H. FM, Editors. 2008, Williams & Wilkins.
  • LH, M. and F. LR, Stabilization of critically ill children, in Nelson Textbook of pediatrics B. RE, K. RM, and J. HB, Editors. 2004, Saunders; 17th edition.
  • Group, A.L.S., Recognition Of The seriously ill child, in Advanced Pediatric Life Support The Practical Approach 2001, BMJ Books 3rd edition.
  • N, E. and S. I., Cardiovascular compromise in the newborn infant, in Avery’s Diseases of the Newborn 8th edn, T. HW, B. RA, and G. CA, Editors. 2005, Elsevier.
  • Association, A.H., Pediatric Advanced Life Support Providers Manual 2011.
  • Organization, W.H., Module 3: Circulation, in Emergency triage assessment and treatment (ETAT) Manual for participants A facilitator guide2005.
Birincil Dil en
Konular Mühendislik ve Temel Bilimler
Dergi Bölümü Makaleler
Yazarlar

Orcid: 0000-0003-0345-6989
Yazar: Ufuk Bal (Sorumlu Yazar)
Kurum: Department of Electrical and Electronics Eng., MSKU
Ülke: Turkey


Bibtex @araştırma makalesi { muglajsci379126, journal = {Mugla Journal of Science and Technology}, issn = {2149-3596}, address = {Mugla Sitki Kocman University}, year = {2018}, volume = {4}, pages = {37 - 40}, doi = {10.22531/muglajsci.379126}, title = {CAPILLARY REFILL TIME MEASUREMENT DEVICE}, key = {cite}, author = {Bal, Ufuk} }
APA Bal, U . (2018). CAPILLARY REFILL TIME MEASUREMENT DEVICE. Mugla Journal of Science and Technology, 4 (1), 37-40. DOI: 10.22531/muglajsci.379126
MLA Bal, U . "CAPILLARY REFILL TIME MEASUREMENT DEVICE". Mugla Journal of Science and Technology 4 (2018): 37-40 <http://dergipark.gov.tr/muglajsci/issue/37235/379126>
Chicago Bal, U . "CAPILLARY REFILL TIME MEASUREMENT DEVICE". Mugla Journal of Science and Technology 4 (2018): 37-40
RIS TY - JOUR T1 - CAPILLARY REFILL TIME MEASUREMENT DEVICE AU - Ufuk Bal Y1 - 2018 PY - 2018 N1 - doi: 10.22531/muglajsci.379126 DO - 10.22531/muglajsci.379126 T2 - Mugla Journal of Science and Technology JF - Journal JO - JOR SP - 37 EP - 40 VL - 4 IS - 1 SN - 2149-3596- M3 - doi: 10.22531/muglajsci.379126 UR - http://dx.doi.org/10.22531/muglajsci.379126 Y2 - 2018 ER -
EndNote %0 Mugla Journal of Science and Technology CAPILLARY REFILL TIME MEASUREMENT DEVICE %A Ufuk Bal %T CAPILLARY REFILL TIME MEASUREMENT DEVICE %D 2018 %J Mugla Journal of Science and Technology %P 2149-3596- %V 4 %N 1 %R doi: 10.22531/muglajsci.379126 %U 10.22531/muglajsci.379126
ISNAD Bal, Ufuk . "KILCAL DAMAR DOLUM SÜRESİ ÖLÇÜM CİHAZI". Mugla Journal of Science and Technology 4 / 1 (Haziran 2018): 37-40. http://dx.doi.org/10.22531/muglajsci.379126